We were faced with a similar situation at our lab.  In our case, the Fire Marshal was who objected to the use of cylinders next to the instrument.  We looked into the remote cylinder idea and the cost to install the double walled lines to travel the extra 100’ or so was over $15000.  So we chose to purchase a hydrogen generator and a monitoring system.

 

 

Kurt France

Technical Manager

Brenntag Pacific Portland

5700 NW Front Ave,

Portland OR 97210

503.248.6265

503.807.2619

 

 

 

From: DCHAS-L Discussion List [mailto:dchas-l**At_Symbol_Here**med.cornell.edu] On Behalf Of Rogers, Janet
Sent: Thursday, October 01, 2015 8:25 AM
To: DCHAS-L**At_Symbol_Here**MED.CORNELL.EDU
Subject: [DCHAS-L] plumbing H2 from remote location

 

To All:

 

We need to use H2 gas for FID-GC.  Our campus EH&S officer wants us to plumb H2 gas from a cylinder placed at some remote location (as yet unknown to me, but I've heard both: a) from the hallway and b) from outside the building) to our instrument lab.  We do not see why we cannot use the H2 cylinder in the instrument lab, next to the instrument, as we have been doing.  Increasing the distance between the H2 cylinder and the instrument seems like it would increase the safety risk. 

 

I'd appreciate hearing your thoughts/comments/experience in handling H2 safely in your labs.

 

 

Janet Rogers, Ph.D.

Professor

Chemistry Department

Edinboro University of Pennsylvania

230 Scotland Road

Edinboro, PA  16444

 

phone: 814.732.1539

e-mail: rogers**At_Symbol_Here**edinboro.edu


NOTICE OF CONFIDENTIAL COMMUNICATION This communication and any subsequent electronic communication relating to this matter, including any attachments, constitute an “electronic communication” within the meaning of the Electronic Communications Privacy Act, 18 U.S.C.A. §2510 or any similar law of any other jurisdiction. In addition, this and any subsequent electronic communication, and any attachments, may be subject to the attorney-client privilege or be considered attorney work product.

This communication, and any subsequent electronic communication relating to this matter, including any attachments, is CONFIDENTIAL and is intended solely for the addressee(s). If you are not an intended recipient, any use, review, dissemination, or copying of any such communication by you is strictly prohibited. If you believe you have received this or any other electronic communication in error, please do not read, copy, use or disclose it to others; also please notify the sender by replying to the communication, and then delete it and all accompanying files from your system. Please verify that you will delete the communication in your reply.

Please note that any views or opinions presented in this communication or any subsequent communication, including any attachments, are solely those of the author or sender and do not necessarily represent those of Brenntag or its affiliated companies. Neither Brenntag nor any of its affiliated companies accepts any liability for any damage caused by any virus transmitted by any email or its attachments.

Previous post   |  Top of Page   |   Next post



The content of this page reflects the personal opinion(s) of the author(s) only, not the American Chemical Society, ILPI, Safety Emporium, or any other party. Use of any information on this page is at the reader's own risk. Unauthorized reproduction of these materials is prohibited. Send questions/comments about the archive to secretary@dchas.org.
The maintenance and hosting of the DCHAS-L archive is provided through the generous support of Safety Emporium.